You are here: Home > HOSPITAL EQUIPMENT>Baby Incubator>Baby Radiant Warmer

Baby Radiant Warmer

Item No.: TRRW-90
Min. Order: 1 Piece

Feature:

▪ Radiant Warmer, Warmer system, infant access (portable).

▪ Mobile freestanding overhead radiant warmer. (Transport, Installation, servicing/maintanence, user level orientation)

▪ Microcontroller based, open type radiant warmer with manual and servo options.

▪ It has facility to display skin set, skin observed temperature in 0 and heat power separately.

▪ Have user friendly touch control panel LED digital display 8 inch

▪ Ceramic or quartz infrared or calrod heater

▪ Audiovisual alarm facility for overheating beyond set temperature range, probe failure, power failure, system failure and heater failure

▪ Machine should sense the skin probe failure and cut off the heater.

▪ Warmer head should rotate and swivel in different direction.

▪ The height of the warmer can adjustable.

▪ X-Ray tray facility

▪ Storage drawer facility

▪ Control Panel should be liquid proof and allow easy and hygienic disinfection

 

Specification:

Observation Light: Variable with intensity range of 0 to 1000 lux(max)or wider.

APGAR Timer 10 second beep at 1 minute, 5 minutes, 10 minutes and 20 minutes.

Skin Temperature Display Range: 18.0°to 45.0°

Resolution: 0.1°C

Accuracy: 0.3

Skin Temperature Setting Range: 33.0° to 38.0°(in 0.1 increments)

Heater Output Indication Numeric display: 0% to 100%, resolution 5%

Dimensions Main body :  850(W)x1170 (D)x(1800-2000)(H) mm (approx.)

Mattress : 470(W)x670(D)x23-30(H)mm approx

Bed Load Capacity: Infant weight up to 10kg

The size of the drop down sides such that it is 5"(°1”)above the mattress surface at least 5 mm thick, clear and transparent.

Bed about 80-100 cm from the floor and 80-90 cm from the heat source.

The baby bed crevice free for ease of cleaning, infection control and mattress used should be of biocompatible material

Heater Element: 500-700 watts.

Bed tilt mechanism: ±5° to 12°

Probe wire is pliable, thin and soft. The attachment site of the probe with the wire should also be pliable and non-stiff.

It has castor wheels  (2 of the castors should have breaks/lock)

Power Supply: Power input 220-240 VAC, 50 Hz fitted with standard plug.